Description
The Agilent 1200 Series variable wavelength detector is designed for highest optical performance, GLP compliance and easy maintenance with:
- higher datarate(27/55Hz) for fast-HPLC with G1314CVWD-SL
- deuterium lamp for highest intensity and lowest detection limit over a wavelength range of 190 to 600 nm,
- optional flow-cell cartridges (standard 10 mm 14 μl, high pressure 10 mm 14 μl, micro 5 mm 1 μl, semi-micro 6 mm 5 μl) are available and can be used depending on the application needs,
- easy front access to lamp and flow cell for fast replacement, and
- built-in holmium oxide filter for fast wavelength accuracy verification.
Performance specifications
Detection type | Double-beam photometer |
Light source | Deuterium lamp |
Wavelength range | 190–600 nm |
Short term noise (ASTM) | ± 0.75 × 10-5 AU at 254 nm |
Drift | 3 × 10-4 AU/hr at 254 nm |
Linearity | > 2 AU (5%) upper limit |
Wavelength accuracy | ±1nm |
Band width | 6.5 nm typical |
Flow cells | Standard: 14-μl volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum High pressure: 14-μl volume, 10-mm cell path length and 400 bar (5880 psi) pressure maximum Micro: 1-μl volume, 5-mm cell path length and 40 bar (588 psi) pressure maximum Semi-micro: 5-μl volume, 6-mm cell path length and 40 bar (588 psi) pressure maximum |
Control and data evaluation | Agilent ChemStation for LC |
Analog outputs | Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output |
Communications | Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional |
Safety and maintenance | Extensive diagnostics, error detection and display (through control module and Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas. |
| |
GLP features | Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter. |
Agilent G1314B - Series 1200
Condition: refurbished
Manufacterer: Agilent Technologies
Software: PC – Chemstation
This refurbished Agilent 1200 VWD is in great shape. It has been tested on working according factory settings by certified service engineers.
The instrument is cleaned (inside and outside) and decontaminated. Broken and spare parts are replaced by original parts of the manufacturer. It comes with warranty. A longer guarantee period is available.