Hoofdafbeelding Agilent G1314C VWD 1200 HPLC series

Agilent G1314C VWD 1200 HPLC series

Article nr.: 8027
Price on request
Reliable
Flexible setup for different analyzes
User Friendly
Software friendly

Description

The Agilent 1200 Series variable wavelength detector is designed for highest optical performance, GLP compliance and easy maintenance with:

  • higher datarate(27/55Hz) for fast-HPLC with G1314CVWD-SL
  • deuterium lamp for highest intensity and lowest detection limit over a wavelength range of 190 to 600 nm,
  • optional flow-cell cartridges (standard 10 mm 14 μl, high pressure 10 mm 14 μl, micro 5 mm 1 μl, semi-micro 6 mm 5 μl) are available and can be used depending on the application needs,
  • easy front access to lamp and flow cell for fast replacement, and
  • built-in holmium oxide filter for fast wavelength accuracy verification.

Performance specifications

Detection typeDouble-beam photometer
Light sourceDeuterium lamp
Wavelength range190–600 nm
Short term noise (ASTM)± 0.75 × 10-5 AU at 254 nm
Drift3 × 10-4 AU/hr at 254 nm
Linearity> 2 AU (5%) upper limit
Wavelength accuracy±1nm
Band width6.5 nm typical
Flow cells

Standard: 14-μl volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum High pressure: 14-μl volume, 10-mm cell path length and 400 bar (5880 psi) pressure maximum

Micro: 1-μl volume, 5-mm cell path length and 40 bar (588 psi) pressure maximum Semi-micro: 5-μl volume, 6-mm cell path length and 40 bar (588 psi) pressure maximum

Control and data evaluationAgilent ChemStation for LC
Analog outputsRecorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output
CommunicationsController-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional
Safety and maintenanceExtensive diagnostics, error detection and display (through control module and
Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
  
GLP featuresEarly maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter.

Agilent G1314B - Series 1200

Condition: refurbished
Manufacterer: Agilent Technologies
Software: PC – Chemstation

This refurbished Agilent 1200 VWD is in great shape. It has been tested on working according factory settings by certified service engineers. 

The instrument is cleaned (inside and outside) and decontaminated. Broken and spare parts are replaced by original parts of the manufacturer. It comes with warranty. A longer guarantee period is available. 

 

Want to stay up to date with the latest news?
Then sign up for our newsletter!